Product Specification
|
Description |
| pH Range |
-2.00 to 16.00 pH |
| Resolution & Accuracy |
0.01 pH & ±0.01 pH |
| mV Range |
-1000 to 1000 mV / 0 to 100% (for -100 to 100 mV) |
| Resolution & Accuracy |
1 mV / 0.1% & ±1 mV / ±0.2% |
| Temperature Range |
-10.0 to 110 °C (14.0 to 230.0 °F) |
| Resolution & Relative Accuracy |
0.1 °C / °F &
±0.5 °C / ±1.0 °F |
| Temperature Sensor |
NTC 300; 2 wire |
| Temperature Compensation |
Auto / Manual
(Independent Process / CAL Temperature) |
| Set Point and Controller Functions |
| Set Point 1 (SP1) / Set Point 2 (SP2) |
-2.00 to 16.00 pH or 0 to 100% or
-1000 to 1000 mV |
| Hysteresis |
0.1 to 1 pH or 1 to 10.0% / 10 to 100 mV |
| Function (Switchable) |
P/PI control (Pulse Length / Pulse Frequency);
Limit Control; Off |
| Proportional Range |
10 to 500% |
| Integral Action Time (IAT) |
0 to 999.9 minutes |
| Adjustable Period with Pulse Length Controller |
0.5 to 20 seconds |
Adjustable Period with Pulse Frequency Controller
|
60 to 120 pulses/min |
| Pickup / Dropout Delay |
0 to 2000 seconds |
| Wash Cycle |
0.1 to 200.0 hours |
| Wash Duration |
1 to 2000 seconds |
Contact Outputs, Controller
|
1 SPDT, 3 SPST relays |
| Switching Voltage / Current / Power |
Max 250 VAC / Max 3A / Max 600VA |
| Alarm Functions |
Function (Switchable)
|
Steady or Fleet (Pulse) |
| Pickup Delay |
0 to 2000 seconds |
| Display |
LCD
|
UV coat, backlit 14 segments display with symbols for status information |
| Backlight |
On/Off selectable with four levels of brightness control |
| Electromagnetic Compliance (EMC) Specifications |
|
Emitted Interference
|
According to EN 61326
|
| Immunity to Interference |
According to EN 61326 |
|
Environmental Conditions
|
| Ambient Temperature Operating Range |
0 to 40 °C |
| Maximum Relative Humidity |
80% up to 31 °C decreasing linearly to 50%
at 40 °C |
| Power Supply |
Input
|
80 to 250 VAC/DC 50/60 Hz Approx. 10 VA |
| Main Fuse |
315 mA time delay, 250 V,
Bussmann BK/GDC-315 mA |
| Pollution Degree |
2 |
| Transient Overvoltage Category |
II |
| Electrical Data and Connections |
| Transmitter Function |
Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated |
| CU 22 Function |
22 mA current output |
| Isolated Output Voltage |
+12V ±0.5 V (Max 50 mA) |
Hold Function Switch
|
To freeze output current and deactivate control relays |
| Load |
Max 600 W |
| Differential pH/ORP Input |
2 pin terminal |
| Connection Terminal |
5-pin, 8-pin, 9-pin and 13-pin terminal, detachable blocks |
| Mechanical Specifications |
| Dimensions (W x H x D) |
144 x 144 x 110 mm |
| Weight |
950g (unit) / 1100g (Packed) |
| Environmental Rating |
IP65 (NEMA 4X) |